1. Products
  2. Medical Adapter
Astek catch your demand on Power Supply Astek catch Your Demand on Power Supplies '
Home Products Medical Adapter AGM018 18W Medical Desktop Power Adapter

AGM018 18W Medical Desktop Power Adapter

*Input 90~264VAC *Output 5~58V*DOE VI*Pass LPS *Protection:Over current/Over voltage/Short Circuit *Means of Patient Protection Design *Possess risk analysis report * UL/cUL60601-1,IEC/EN/ANSI/AAMI ES 60601-1 (ED 3.1),EMC: IEC60601-1-2:2014 (ED 4.0),Means of Protection: 2 X MOPP *Two years warranty

187150038a367d9aad0add8b0f23c371.jpg

AGM018  18W Medical Desktop Adapter battery

*Universal AC Input Range

*DOE VI

*Pass LPS

*Protections: Over current/ Over voltage/Short Circuit

*Means of Patient Protection Design

*Possess risk analysis report

* UL/cUL60601-1,IEC/EN/ANSI/AAMI ES 60601-1 (ED 3.1)

*EMC: IEC60601-1-2:2014 (ED 4.0)
*Means of Protection: 2 X MOPP

*Warranty two Years

*Application: Medical power, Medical charger(CV mode), Power Supply for Health care

Input Voltage: 90~264VAC

Frequency: 50~60Hz

Input current: 0.45-0.2A max.   

Leakage Current: 0.1mA max. @240VAC /50Hz                          

Inrush current: 50A @115VAC; 90A @230VAC at 25°C cold start  

Part No          Rated O/P(current max.)        Ripple & Noise(mV p-p)

AGM018-S05-z      5-6V/3-2.5/15W                         80

AGM018-S75-z      6-8V/2.5-1.87A/15W                  80

AGM018-S09-z      8-11V/2.25-1.64A/18W            100

AGM018-S12-z      11-13V/1.64-1.38A/18W          120

AGM018-S15-z      13-16V/1.38-1.13A/18W          150

AGM018-S18-z      16-21V/1.13-0.86A/18W          150

AGM018-S24-z      21-27V/0.86-0.67A/18W          150

AGM018-S27-z      27-33V/0.67-0.55A/18W          240

AGM018-S48-z      33-58V/0.55-0.32A/18W          240

z: C6/C14/C8 inlet, weight:Appro.140-150g   
Attached: AGM018 desktop mechanical derating curve
  
Load Regulation: +/-5% Max.

Line Regulation: +/-1% Max.  

Insulation resistance: 50Mohm min. from Input to output                    

Over Current: auto-recovery

Over Voltage Protection: Protected by Zerner diode

Short circuit Protection: Hiccup mode; auto recovery

Setup/Hold up/ Rise time: Typ.: 1000mS/10mS/50mS at full load

Efficiency : DOE VI

Derating : Derated from 100% at +40°C linearly to 50% at +70°C

Temperature Coefficient: All output +/-0.04% / °C max.

Withstand Voltage: 5656VDC from input to output

Operating Temperature: 0~40degree C

Storage Temperature: -40~85degree C

Storage Humidity: 5~95% RH

Transient Response: Max. Excursion of 4% better on all models. Recovering to 1% of final value within

                 500uS. after a 25% step load change

Safety: UL/cUL60601-1,IEC/EN/ANSI/AAMI ES 60601-1 (ED 3.1)

EMC: IEC60601-1-2:2014 (ED 4.0)