Home
1
Products
2
AGM006 5W-6W Medical wall mount Adapter 3
Astek catch your demand on Power Supply Astek catch Your Demand on Power Supplies '
Home Products Medical Adapter AGM006 5W-6W Medical wall mount Adapter
136

AGM006 5W-6W Medical wall mount Adapter

*Input 90~264VAC, Output 5V~24V *No load power consumption : <0.3W *DOE VI*Protections: Overcurrent/ Overvoltage/Short Circuit *Means of Patient Protection Design *Possess risk analysis *Two years warranty *UL/cUL60601-1,IEC/EN/ANSI/AAMI ES 60601-1 (ED 3.1), EMC: IEC60601-1-2:2014 (ED 4.0), Means of Protection: 2 X MOPP

4da217d938c7b3d710b6a8a3b1efb2d2.jpg

AGM006  5W-6W Medical wall mount Adapter

*Universal AC Input Range

*No load power consumption: =8m/Second min.

*DOE VI

*Protections: Overcurrent/ Overvoltage/Short Circuit

*Means of Patient Protection Design

*Possess risk analysis report

* UL/cUL60601-1,IEC/EN/ANSI/AAMI ES 60601-1 (ED 3.1)

*EMC: IEC60601-1-2:2014 (ED 4.0)
*Means of Protection: 2 X MOPP

*Warranty two Years

*Application: Medical power, Medical charger(CV mode), Power Supply for Health care

Input Voltage: 90~264VAC

Frequency: 50~60Hz

Input current: 0.18-0.09A   

Leakage Current: 0.1mA max.                           

Inrush current: 30A @115VAC; 60A @230VAC at 25°C cold start   

Part No          Rated O/P(current max.)                                 Ripple & Noise(mV p-p)

AGM006-S05-z      5-6V/1.2-1.0A/5-6W                     80

AGM006-S75-z      6-8V/1.0-0.75A/6W                       80

AGM006-S09-z      8-11V/0.75-0.55A/6W                   80

AGM006-S12-z      11-13V/0.55-0.46A/6W               120

AGM006-S15-z      13-16V/0.46-0.38A/6W               150

AGM006-S18-z      16-21V/0.38-0.29A/6W               180

AGM006-S24-z      21-24V/0.29-0.26A/6W               200

z: US/EU, 2pin & 3pin AU plug, UK plug, weight: 70~100g

Load Regulation: +/-5% Max.

Line Regulation: +/-1% Max.  

Insulation resistance: 50Mohm min. from Input to output                    

Over Current/ Over Voltage Protection: Auto Recovery

Hold Up Time: >=8m/Second min.

Efficiency: DOE VI

Derating : Derated from 100% at +40°C linearly to 50% at +70°C

Temperature Coefficient: All output +/-0.04% / °C max.

Withstand Voltage: 5656VDC from input to output

Operating Temperature: 0~40degree C

Storage Temperature: -40~85degree C

Storage Humidity: 5~95% RH

Transient Response: Max. Excursion of 5% better on all models. Recovering to 3% of final value within 500n/sec. after a 50% step load change

Safety: UL/cUL60601-1,IEC/EN/ANSI/AAMI ES 60601-1 (ED 3.1)

EMC: IEC60601-1-2:2014 (ED 4.0)

Spec Attached: AGM006.docx