AGM050 50W Medical Desktop Adapter
*Medical adapter *Input 90~264VAC *Output 3V~58V*High Efficiency and reliability*Protections: Overcurrent/Overvoltage/Over Temperature* UL/cUL60601-1, IEC/EN/ANSI/AAMI ES 60601-1 (ED 3.1), EMC: IEC60601-1-2:2014 (ED 4.0), CE, FCC, Means of Protection: 2 X MOPP, *Warranty 2years
AGM050 50W Medical Desktop Adapter
*Universal AC Input Range
*High Efficiency and reliability
*Protections: Overcurrent/ Overvoltage/Over Temperature
* UL/cUL60601-1, IEC/EN/ANSI/AAMI ES 60601-1 (ED 3.1)
* EMC: IEC60601-1-2:2014 (ED 4.0), CE, FCC
* Means of Protection: 2 X MOPP
*Warranty two Years
*Application: Medical power, Medical charger(CV mode), Power Supply for Health care
Input Voltage: 90~264VAC
Frequency: 50~60Hz
Input current: 1.5-0.8A max.
Leakage Current: 0.3mA max. @240VAC /50Hz
Inrush current: 65A @115VAC; 90A @230VAC at 25°C cold start
Part No Rated O/P(current max.) Ripple & Noise(mV p-p)
AGM050-S03-z 3-5V/7.5A/37.5W 50
AGM050-S05-z 5-6V/7.5-6.25A/37.5W 50
AGM050-S75-z 6-8V/6.66-5A/40W 75
AGM050-S09-z 8-11V/5.62-4.09A/45W 90
AGM050-S12-z 11-13V/4.54-3.84A/50W 100
AGM050-S15-z 13-16V/3.84-3.12A/50W 100
AGM050-S18-z 16-21V/3.12-2.38A/50W 150
AGM050-S24-z 21-27V/2.38-1.85A/50W 150
AGM050-S27-z 27-33V/1.85-1.51A/50W 200
AGM050-S48-z 33-58V/1.51-1.04A/50W 200
z: C6/C14 inlet
Mechanical Drawing
Attached: AGM050 desktop mechanical
Load Regulation: +/-5% Max.
Line Regulation: +/-1% Max.
Insulation resistance: 50Mohm min. from Input to output
Over Current: All output protected to short circuit conditions
Over Voltage Protection: Provided on outputs set at 112-132% of its nominal output voltage
Over temperature Protection: 130°C Activated
Setup/Hold up/ Rise time: Typ.: 1000mS/10mS/50mS at full load
Derating : Derated from 100% at +40°C linearly to 50% at +70°C
Temperature Coefficient: All output +/-0.04% / °C max.
Withstand Voltage: 5656VDC from input to output
Operating Temperature: 0~40degree C
Storage Temperature: -40~85degree C
Storage Humidity: 5~95% RH
Transient Response: Max. Excursion of 4% better on all models. Recovering to 1% of final value within
500uS. after a 25% step load change
Safety: UL/cUL60601-1, IEC/EN/ANSI/AAMI ES 60601-1 (ED 3.1)
EMC: IEC60601-1-2:2014 (ED 4.0), CE, FCC
Spec Attached: AGM050 desktop.docx